P309 – Exit Slit Unit SXR LCLS SLAC

Application

Exit slit unit for soft X-ray FEL radiation

Year of delivery

2010

Installation site

Soft X-ray beamline at LCLS, Stanford Linear accelerator Center, USA

Design Features

Performance Features

  • Precision and slit opening repeatability smaller than 100 nm.
  • Parallelism smaller than 4 µm over 10 mm edge length.
  • Straightness smaller than 0.6 µm over 10 mm edge length.
  • Very small deviation (submicrometer) of the slit opening calibration curve from the theoretical non-linear slit opening function.

Outer Dimensions

P309 - Outer dimensions - Exit slit unit for soft X-ray FEL radiation at Soft X-ray beamline at LCLS, Stanford Linear accelerator Center, USA - Bestec GmbH

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, horizontal axis

Main flanges

DN160CF cluster flanges

Material

1.4301

Base pressure

10-10 mbar range

Vertically defining slit (VDS)

Slit blades

Material

B4C

Straightness

< 0.60 µm over 10 mm (1σ)

Parallelism

0.37 µm /mm (85 µm slit opening)

Slit opening

Range

0 µm to 10 mm

Repeatability

± 0.043 µm up to 350 µm slit opening

Resolution (design value)

0.04 µm /motor fine step
(1/128, recommended)
0.05 µm /encoer count

Performance Test Results

Slit width calibration (VDS)
P309 - Slit width calibration (VDS) - Exit slit unit for soft X-ray FEL radiation at Soft X-ray beamline at LCLS, Stanford Linear accelerator Center, USA - Bestec GmbH
Slit opening repeatability (VDS)
P309 - Slit opening repeatability (VDS) - Exit slit unit for soft X-ray FEL radiation at Soft X-ray beamline at LCLS, Stanford Linear accelerator Center, USA - Bestec GmbH
Slit blade parallelism (VDS)
P309 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray FEL radiation at Soft X-ray beamline at LCLS, Stanford Linear accelerator Center, USA - Bestec GmbH