P506 – Exit Slit Unit HINESS HEPS IHEP CAS

Application

Exit slit unit for soft X-ray radiation

Year of delivery

2024

Installation site

HINESS beamline at High Energy Photon Source, Institute of High Energy Physics,  China

Design Features

  • Double slit mechanism – vertically and horizontally defining slits.

  • Motorized and encoded slits opening, symmetric opening, non-linear opening characteristics.
  • OFHC slit blades.
  • Electrical isolation of horizontally defining slit blades.

  • Isolated bias plate with fluorescing screen.
  • Manually retractable photodiode.
  • EtherCAT based motion control system, synchronized encoder readout, software with server-client architecture, TCP-IP protocol, EPICS interface.

  • Adjustable support structure.

Performance Features

  • Precision and slit opening repeatability smaller than 150 nm.
  • Parallelism smaller than 0.55 µm over 10 mm edge length.
  • Straightness smaller than 0.2 µm over 10 mm edge length.
  • Very small deviation of the slit opening calibration curve from the theoretical non-linear slit opening function.

Outer Dimensions

P506 - Outer dimensions - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, horizontal axis

Main flanges

DN160CF cluster flanges

Material

1.4301

Base pressure

10-10 mbar range

Horizontally defining slit (HDS)

Slit blades

Material

OFHC

Straightness

< 0.50 µm over 25 mm (1σ)

Parallelism

-0.01 μm /mm
(ca. 50 μm slit opening)
0.002 μm /mm
(ca. 100 μm slit opening)
0.05 μm /mm
(ca. 150 μm slit opening)

Slit opening

Range

1.70 µm to 4.10 mm

Repeatability

0.10 μm at ca. 50 μm slit opening
0.09 μm at ca. 100 μm slit opening
0.08 μm at ca. 150 μm slit opening

Resolution (design value)

0.00156 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Vertically defining slit (VDS)

Slit blades

Material

OFHC

Straightness

< 0.30 µm over 30 mm (1σ)

Parallelism

-0.04 μm / mm
(ca. 53 μm slit opening)
-0.02 μm /mm
(ca. 152 μm slit opening)

Slit opening

Range

2 µm to 4 mm

Repeatability

0.04 μm at ca. 23 μm slit opening
0.06 μm at ca. 52 μm slit opening
0.04 μm at ca. 151 μm slit opening

Resolution (design value)

0.00156 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Performance Test Results

Slit width calibration (HDS)
P506 - Slit width calibration (HDS) - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH
Slit opening repeatability (HDS)
P506 - Slit opening repeatability (HDS) - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH
Slit blades parallelism (HDS)
P506 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH
Slit  blades parallelism (HDS)
P506 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH
Slit blades parallelism (HDS)
P506 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH
Slit width calibration (VDS)
P506 - Slit width calibration (VDS) - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH
Slit opening repeatability (VDS)
P506 - Slit opening repeatability (VDS) - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH
Slit blades parallelism (VDS)
P506 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH
Slit blades parallelism (VDS)
P506 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH
Residual gas analysis
P506 - Residual gas analysis - Exit slit unit for soft X-ray synchrotron radiation, HINESS beamline at High Energy Photon Source, Institute of High Energy Physics, Chinese Academy of Science - Bestec GmbH